Atlas Scientific dissolved oxygen (DO) probes are galvanic probes that utlize a membrane. When removing the cap for the probe, be sure to pull (and not twist) the cap off.
- Open the Atlas Scientific DO EZO calibration documentation to follow along
- Remove the cap of the DO probe and place on the counter
- Type
r
(without quotes) into the serial monitor and press enter to prompt the probe to collect a reading. The wordsSuccess
and 2 values should appear in the serial monitor. Continuously repeat this for 5-30 seconds until values stabilize. - Type
cal
into the serial monitor, and pressenter
. This calibrates the probe to full DO saturation using the air
Two-point calibration (for readings < 1 mg/L)
- Place DO probe in the Zero DO calibration solution and stir the probe to remove trapped air. Wait between 30 and 90 seconds while typing
r
and pressingenter
continuously. - Once values stabilize, type
cal,0
and pressenter
. Values from the probe (using ther
command) should now show zero.